Operators manual – Watson-Marlow PF22 User Manual

Page 17

Advertising
background image

Operators Manual

PF22

PF22 OM EN 74-132-005 v1.02.doc

Version: 1.02

Page 17 of 34


46. Select language
Value: An integral number.
This version offers a choice of 2 languages.

1 - English
2 - German

The languages will be active on the display and on the print-outs.

47. Printer set-up
Function 47 tells the system which protocol to use when transmitting to connected printer.
When this function is activated, it will first require the operator to enter the transmission velocity.
One of the following values must be chosen:

75

110

134

150

300

600

1200 1800 2000 2400 4800 9600

Subsequently PF22 will ask the operator to enter a protocol number. There is a choice between
the following protocols:

1 :

7 data bits

1 stop bit

no parity

2 :

7 data bits

1 stop bit

even parity

3 :

7 data bits

1 stop bit

uneven parity

4 :

7 data bits

2 stop bits

no parity

5 :

7 data bits

2 stop bits

even parity

6 :

7 data bits

2 stop bits

uneven parity

7 :

8 data bits

1 stop bit

no parity

8 :

8 data bits

1 stop bit

even parity

9 :

8 data bits

1 stop bit

uneven parity

10 :

8 data bits

2 stop bits

no parity

11 :

8 data bits

2 stop bits

even parity

12 :

8 data bits

2 stop bits

uneven parity


72. Volume format

This function sets the unit for volume. If the value 1 is chosen then the volume unit is set to ml and
if value 2 is chosen then the volume unit is set to gram. The unit ml or gram is shown in F1 and log
print outs.

80. Reset memory
This function will reset the memory with the exception of the part used for saving programs.
PF22 will pass into stand-by, and when switched on again the built-in parameters will be valid.

IF PF22 DOES NOT WORK OR DOES NOT OPERATE AS EXPECTED,

ACTIVATE FUNCTION 80.

Advertising