2-6 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT

ADAPTOR PLATE NOT FURNISHED BY ROSEMOUNT.

INSTALLATION FOR MASONRY

WALL STACK CONSTRUCTION

INSTALLATION FOR METAL

WALL STACK OR DUCT

CONSTRUCTION

13 (0.50)

95 (3.75)

MIN DIA HOLE

IN WALL

STACK OR DUCT

METAL WALL

MTG HOLES

SHOWN ROTATED

45 OUT OF

TRUE POSITION

o

WELD OR BOLT MOUNTING

PLATE TO METAL WALL

OF STACK OR DUCT.

JOINT MUST BE AIRTIGHT.

13 (0.50)

114 (4.50)

O.D. REF

PIPE 4.00 IN. SCHED 40

PIPE SLEEVE (NOT

BY ROSEMOUNT)

LENGTH BY CUSTOMER

MASONRY

STACK WALL

OUTSIDE WALL

SURFACE

JOINT MUST

BE AIRTIGHT

MTG HOLES

SHOWN ROTATED

45 OUT OF

TRUE POSITION

o

FIELD WELD

PIPE TO

MOUNTING PLATE

BOLT MOUNTING

PLATE TO OUTSIDE

WALL SURFACE

NOTE: DIMENSIONS IN MILLIMETERS WITH

INCHES IN PARENTHESES.

82.5 (3.25)

MIN DIA HOLE

IN WALL

STACK OR DUCT

METAL WALL

WELD OR BOLT MOUNTING

PLATE TO METAL WALL

OF STACK OR DUCT.

JOINT MUST BE AIRTIGHT.

FIELD WELD

PIPE TO

MOUNTING PLATE

102 (4.0)

O.D. REF

PIPE 3.5 IN. SCHED 40

PIPE SLEEVE (NOT

BY ROSEMOUNT)

LENGTH BY CUSTOMER

MASONRY

STACK WALL

OUTSIDE WALL

SURFACE

JOINT MUST

BE AIRTIGHT

BOLT MOUNTING

PLATE TO OUTSIDE

WALL SURFACE

PROBE MOUNTING

ABRASIVE SHIELD MOUNTING

27540004

Figure 2-5. Hazardous Area Oxymitter 4000 Mounting Plate Installation