With remote electronics -12, Hazardous area oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 User Manual

Page 58: Wiring diagram

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1
July 2004

2-12 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

TYPE K
THERMOCOUPLE
SIGNAL

HEATER POWER
(BELOW COVER)

OXYGEN

SIGNAL

+

+

-

-

TERMINAL
BLOCK

GROUND

LUGS

37270023

LINE VOLTAGE

4-20 mA

SIGNAL

A

C

L1

A

C

N

+

+

-

-

4-20

TERMINAL
BLOCK

GROUND LUGS

1

2

3

4

GRN

ORN

YEL

RED

BLK

BLK

ORN

RED

BLK

HTR

T/C

O

CELL

2

WIRING DIAGRAM

TO ELECTRONICS

FROM PROBE

WHT

BRN

GRN

RED

BRN

BLU

CALIBRATION
HANDSHAKE/
LOGIC I/O

PROBE

REMOTE

ELECTRONICS

TO IMPS 4000 IF USED.

REFER TO IMPS 4000

INSTRUCTION MANUAL.

LOGIC I/O +
LOGIC I/O -

LINE 1

NEUTRAL

GROUND

90-250 VAC, 50-60 HZ
LINE VOLTAGE INPUT

INTERCONNECTING

CABLE

4-20 mA +
4-20 mA -

GROUND

5

6

7

8

(WHT)

(RED)

(BLU)

(HAZ.)

(BRN)

(GRN)

(BRN)

BLU

YEL

WHT

GRN/YEL

or GRN

G.P

.

HAZARDOUS AREA

OXYMITTER

GENERAL PURPOSE

OXYMITTER

Figure 2-10. Electrical Installation – Hazardous Area Oxymitter 4000

with Remote Electronics (Sheet 1 of 2)

Advertising