5-2 Startup and Operation with Keypad

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

d. Keypad

The five membrane keys on the membrane

keypad are only used during calibration to
adjust the high and low gas and to initiate
the calibration sequence (Figure 5-2).

e. Reference

Air

Ensure reference air, if used, is set to 56.6

l/hr (2 scfh).

5-2 OPERATION

a. Overview

Ensure the Hazardous Area Oxymitter 4000

is at normal operation. The diagnostic LEDs
will display the operating cycle. All other
LEDs should be off (See Figure 5-3).

1. DIAGNOSTIC ALARM LEDS. If there

is an error in the system, one of these
LEDs will flash various blink codes
(see Section 8, TROUBLESHOOT-
ING). In the case of multiple errors,
only one will be displayed based on a
priority system. Correct the problem
and cycle power. The operating display
will return or the next error will be dis-
played. The alarms are:

HEATER T/C
HEATER
O

2

CELL

CALIBRATION

2. CALIBRATION RECOMMENDED

LED. Turns on when the system
determines that a calibration is recom-
mended. Further information is avail-
able in Section 9, MAINTENANCE
AND SERVICE.

3. TEST POINTS. Test points 1 through 6

allow you to monitor with a multimeter:
the heater thermocouple, the O

2

cell

millivolt value, and the process O

2

.

(a) TP1 and TP2 monitor the oxygen

cell millivolt output, which equates
to the percentage of oxygen
present.

(b) TP3 and TP4 monitor the heater

thermocouple.

(c) TP5 and TP6 monitor the process

gas or the calibration gas
parameter.

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

02 CELL mV +
02 CELL mv -
HEATER T/C +
HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% 02

MEMBRANE

KEYS

MEMBRANE

KEYS

MEMBRANE

KEY

DIAGNOSTIC

LEDS

22220023

Figure 5-2. Calibration Keys