Troubleshooting -1, Overview -1, Figure 8-1. o – Emerson Process Management OXYMITTER 4000 User Manual

Page 97: Sensor mv reading vs. % o, At 736°c (reference air, 20.9% o, Ee section 8, t, Section 8, Hazardous area oxymitter 4000, O sensor performance at 736 c

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-1

Hazardous Area Oxymitter 4000

SECTION 8

TROUBLESHOOTING

8-1 OVERVIEW

While the Hazardous Area Oxymitter 4000
electronics provides a significant number of
diagnostic alarms to assist in troubleshooting
potential problems, it’s good to place these
alarms in perspective with respect to the
instrument’s operating principles:

a. When the Zirconium Oxide sensing cell is

heated to its setpoint [736°C (1357°F)], the
cell will generate a voltage that represents

the difference between the process O

2

%

and the reference O

2

% inside the probe

(20.95% O

2

ambient air).

b. Test points, Figure 8-1, are provided to read

the raw millivolt value generated by the
ther-mocouple that controls both the cell
temperature and the raw cell signal.

c. The cell temperature at test points 3 and 4

should always be stable at approximately
29 to 30 millivolts, which represents the
736°C setpoint temperature.

37260043

100

10

1

0.01

0.1

0

0

50

100

150

200

Concentration O (%)

2

EMF (mV)

O Sensor Performance at 736 C

2

O

O

2

%

100

20

15

10

9

8

7

6

5

4

EMF(mV)

-34

1.0

7.25

16.1

18.4

21.1

23.8

27.2

31.2

36.0

O

2

%

3

2

1

0.8

0.6

0.5

0.4

0.2

0.1

0.01

EMF(mV)

42.3

51.1

66.1

71.0

77.5

81.5

86.3

101.4

116.6

166.8

Figure 8-1. O

2

Sensor mV Reading vs. % O

2

at 736°C (Reference Air, 20.9% O

2

)

8

Advertising