Compliance information safety, Electromagnetic compatibility – Precision Digital PD8-765 User Manual

Page 10

Advertising
background image

PD8 Series Explosion-Proof Instrument Supplemental Manual

10

COMPLIANCE INFORMATION

Safety

LOW VOLTAGE
DIRECTIVE

EN 61010-1:2010
Safety requirements for measurement, control, and
laboratory use

Electromagnetic Compatibility

EMISSIONS

EN 55022:2010
Class A ITE emissions requirements

Radiated
Emissions

Class A

AC Mains
Conducted
Emissions

Class A

IMMUNITY

EN 61326-1:2006
Measurement, control, and laboratory equipment

EN 61000-6-2:2005
EMC heavy industrial generic immunity standard

RFI - Amplitude
Modulated

80 -1000 MHz 10 V/m 80% AM (1 kHz)
1.4 - 2.0 GHz 10 V/m 80% AM (1 kHz)
2.0 - 2.7 GHz 10 V/m 80% AM (1 kHz)

Electrical Fast Tran-
sients

±2 kV AC mains, ±1kV other

Electrostatic
Discharge

±4 kV contact, ±8kV air

RFI - Conducted

10 V, 0.15-80 MHz, 1kHz 80% AM

AC Surge

±2 kV Common, ±1kV Differential

Surge

1 kV (CM)

Power-Frequency
Magnetic Field

3 A/m 70%V for 0.5 period

Voltage Dips

40% V for 5 & 50 periods
70% V for 25 periods

Voltage
Interruptions

<5% V for 250 periods

Advertising