Troubleshooting, Troubleshooting tips – Precision Digital PD8-765 User Manual

Page 28

Advertising
background image

PD8 Series Explosion-Proof Instrument Supplemental Manual

28

TROUBLESHOOTING

The ProtEX-MAX rugged design should make it unusual for the installer or operator
to refer to this section of the manual. If the instrument is not working as expected,
refer to the recommendations below.

Troubleshooting Tips

Symptom Check/Action

SafeTouch buttons do not respond

If mechanical button was pushed. The
SafeTouch buttons will be re-enabled
automatically 60 seconds after the last
button push.

If slide switch on connector board is in
DISABLE position, switch to ENABLE.

Strong direct sunlight may interfere with
SafeTouch button operation. It is
recommended to operate the buttons by
standing so as to block direct sunlight.

Serial Communications Power LED
Indicator is off

1. Check modular cable connection

2. Check power to the device

Meter not communicating with
MeterView, MeterView Pro, or other
programs

1. Check serial cable

2. Check protocol selected on device

3. Check instrument address & baud rate

4. Check program address & baud rate

If only the TX (or DATA IN) data status
LED is flashing when serial
communications attempted

1. Check serial cable

2. Check protocol selected on device

3. Check instrument address & baud rate

4. Check program address & baud rate

If both data status LEDs (TX and RX)
are off when trying to communicate

Remove all unnecessary cables and
instruments from the bus. Try getting the
system to work with only one device (to
ease troubleshooting) and then expand
the system one device at a time.

Communications slow

Increase the baud rate

Random communication errors

1.

Increase the TX delay time

2.

Decrease the baud rate

Power LED is off

1.

Check modular cable connection

2.

Check power to instrument

Other symptoms not described above

Call Technical Support for
assistance or the instruction manual
included with this ProtEX-MAX.