To Output Basic Waveform

DG1000Z can output basic waveforms (Sine, Square, Ramp, Pulse and Noise) from

one of the channels separately or from the two channels at the same time. This

section introduces how to output a sine waveform (Frequency: 20kHz, Amplitude:

2.5Vpp, DC Offset: 500mV

DC

, Start Phase: 90°) from the [CH1] connector.

1. To select output channel

Press CH1|CH2 to select CH1. Now the boarder of the channel status bar is
displayed in yellow.

2. To select the Sine

Press Sine to select the sine waveform. The backlight goes on and the

corresponding menu is displayed in the right of the screen.

3. To set the frequency/period

Press Freq/Period to highlight “Freq”, and then use the numeric keyboard to

input 20. Then select kHz from the pop-up menu.
 The frequency ranges from 1μHz to 60MHz.
 The frequency units available are MHz, kHz, Hz, mHz and μHz.
 Press this softkey again to switch to period setting.
 The period units available are sec, msec, μsec and nsec.

4. To set the amplitude

Press Ampl/HiLevel to highlight “Ampl”, and use the numeric keyboard to

input 2.5. Then, select Vpp from the pop-up menu.
 The amplitude range is limited by the impedance and frequency/period.
 The amplitude units available are Vpp, mVpp, Vrms, mVrms and dBm (dBm

is only valid when the setting in UtilityChannel SetOutput Set

Imped is not HighZ).

 Press this softkey again to switch to high level setting.
 The high level units available are V and mV.

5. To set the offset

Press Offset/LoLevel to highlight “Offset”, and then use the numeric keyboard

to input 500. Then, select mV

DC

from the pop-up menu.

 The range of the offset is limited by the impedance and frequency/period.
 The DC offset voltage units available are V

DC

and mV

DC

.

 Press this softkey again to switch to low level setting. The low level must be

lower than the high level at least 1mV (when the output impedance is 50Ω).

 The low level units available are V and mV.