Table 3-19. mbt-g compatibility modes, 10 command language, 3 remote mode programming – KEPCO MBT Series User Manual

Page 88: Command language -50, Remote mode programming -50, 19 mbt-g compatibility modes -50

Advertising
background image

3-50

MBTSVC111609

1.

Press MENU ON/OFF to enter the menu, then NEXT until "Compatible Mode" is dis-

played; press ENTER to select this function.

2.

The display shows either "Compatibility ON" or "Compatibility OFF." Use the

slew keys to toggle between the 2 modes.

3.

Press MENU ON/OFF to return to menu level. Press it again to exit menu system.

3.2.2.10 COMMAND LANGUAGE

The MBT-G allows selection of either SCPI or CIIL command languages for remote program-
ming.

1.

Press MENU ON/OFF to enter the menu, then NEXT until "Command Language" is dis-

played; press ENTER to select this function.

2.

The display shows either "SCPI Programming" or "CIIL Programming." Use the

slew keys to toggle between the 2 modes.

3.

Press MENU ON/OFF to return to menu level. Press it again to exit menu system.

3.3

REMOTE MODE PROGRAMMING

NOTE: The VISA query function (included in the latest versions of the VISA libraries) is not

supported by the MBT Power Supply. In newer XP and Vista computers a delay is
needed to insure the MBT Power Supply has time to return the response to the query.
The VISA query has no provisions to add a delay, nor can it issue a series of read
strobes between sending the request and receiving data from the device. Instead of
the VISA query, use one of the two methods described below. Method 2 is recom-
mended.

1. Use a VISA Write followed by VISA Read with a small delay between the functions.
The delay is determined by computer speed. A 2 millisecond delay between the Write
and Read functions will be sufficient to insure that there is enough time to receive a
response for all valid queries.

2. Follow the VISA Write with a series of VISA Read strobes while waiting for the data
available bit to be true, followed by the VISA Read function when data is available. This
method provides the highest throughput.

TABLE 3-19. MBT-G COMPATIBILITY MODES

Function

MODE 0 (OFF)

MODE 1 (ON)

Device Clear

Does not reset power modules

Resets all power modules

Status

Uses Status Instrument Registers

Does not use Status Instrument Registers

PON enable

All status register enables are set to 0

All status register enables are set to 32767

syst:vers?

returns 1997.0

returns blank string

Advertising