Escort 3016B User Manual
Page 56

7
Mathematical calculations
Page 2
Function: selection of the calculation function made in the selected calculation of #Param
You can choose among 20 different mathematical calculations in 3 categories:
-
Amplitude: min values, max values, peak to peak, base, highest, amplitude, overshoots
-
Time: frequency, period, increasing time, decreasing time, positive widths, negative
widths, positive and negative duty factors
-
Calculation: average values, cyclic average, cyclic RMS and effective RMS
You can display up to 5 simultaneous calculations on screen.
The display is made in rectangles above the diagrams where are also specified:
-
the number of the channel (with its colour)
-
the type of calculation
-
the value of the calculation
In Display function, the calculations are made in real time and the display is updated every 300
ms.
The calculation is made on the 1000 dots displayed on screen. Thus, the time resolution is 0.1 %.
The calculations can concern every channels. However, you cannot assign calculations:
- on the supplementary channels that are functions of other channels
(example F3=A1+B2)
- if the channels have been recorded (validation ON/OFF)
7.2.
Types of calculations
Explicative scheme
Math.
Functions
Calculation
Remarks
Max
High
90 % Ampl
10 % Ampl
Lo
w
Min
Rise time
Falltime
Ampl
P
eak_peak