Escort 3016B User Manual
Page 111
15
Ethernet interface
Page 13
14.5.3.Functions of the channels and between the channels
14.5.4.Changing mode
14.5.5.Current function (page)
HEADER
PARAMETERS
EXAMPLES
THREshold
P1,P2,P3
:THRES S1,ON,10
Defines the thresholds
threshlod S1 is 10 Volts
P1=SI or S2
P2=ON or OFF ( plotting threshold)
P3=Value of threshold
THREshold ?
Returns values of the 2 thresholds of CHANNEL
RDC ?
Returns values of all valid channels (ON) and logic channels value
HEADER
PARAMETERS
EXAMPLES
FUNCMATH
P1
CHAN 2;FUNCTION LOGX;
Select a mathematical function for CHANNEL
channel 2 is aLog(x)+b
P1 = Type of function :
NONe, UNIT, AX, ABSX, SQRX, SQROOTX, LOGX, EXPX
(without, change unit, ax+b, a|x|+b, ax? b, ...)
FUNCMATH ?
Returns the function of CHANNEL
COEFf
P1,P2
:COEF A,2;COEF B,0
Defines the function coefficients
A is 2
P1 is A, B ,C or X1,X2,Y1,Y2
B is zero
COEFf ?
Returns coefficients values of the function of CHANNEL
UNITFunction
P1
UNITF 'DB'
Defines the unit of the function
P1 = name of the unit ( max 6 caracteres) beetween " or '
UNITFunction ?
Returns the function unit name
FUNCXY
P1,P2,P3
Defines the function beetween channels F(X,Y)
P1= first channel ( from A1 to FF)
P2= Operator PLUS,MINUS,MULT,DIV
P3= second channel ( from A1 to FF)
FUNCXY ?
Returns the function
RDUnit
P1
Select the measurement unit :
P1: ISO channels unit
NORM : nomalized unit fron 0 to 10000
FUNCTion
P1
Validity of functions
P1=ON or OFF
FUNCTIon ?
Returns validity of functions
CHAN FB;FUNCXY A1,PLUS,A2
RDU ISO;ONOFF ALL,OFF;ONOFF
A1,ON;ONOFF A3,ON;RDU ISO;DRC ?
HEADER
PARAMETERS
EXAMPLES
MODE
P1
Defines the mode of using the instrument
P1=DIRect, MEMory, FILE, GONOgo
MODE ?
Returns current mode
HEADER
PARAMETERS
EXAMPLES
PAGe
P1
CHAN A3;SCREEN CHAN
Display a screen on the LCD
Setup of channel A3
P1 =
SETUP : setup
CHAN : channel setup of CHANNEL
TRigger : triggering
CHArt: thermal paper setup
SCOpe : direct display
REPLay : replay memory display