gohastings.com POWERPOD 400 User Manual

Page 20

Advertising
background image

Page 20 of 42

10.2.3. Selecting Unit-of-Measure





After entering the SETUP/CAL mode and after selecting a channel number, the SETUP/CAL menu
is displayed. Choose DISPLAY (1) + ENTER and then select UNITS by pressing the number “2” key +

ENTER. The UNITS menu will be displayed allowing the selection of over 50 different units of measure
(UOM) simply by pressing the number key corresponding to the desired UOM. The number can be

found by scrolling up or down until the desired UOM is displayed or by consulting Appendix B, at the end

of this manual. Once selected, the UOM will be displayed in character spaces 10 to 14 on the appropriate
line. Selecting UOM = “0” results in the UOM characters being blanked.

10.2.4. Selecting Gas ID





While the SETUP/CAL menu is displayed, Select the DISPLAY menu by pressing 1 + ENTER. Then

choose GASID by pressing 3 + ENTER. The GASID menu will be displayed allowing the selection of
nearly one hundred different chemical symbols simply by pressing the number key corresponding to the

desired gas. The number can be found by scrolling up or down until the desired gas is displayed or by

consulting Appendix C at the end of this manual. Once selected, the formula will be displayed in
character spaces 16 through 20 on the appropriate line.

10.2.5. Setting A/D Conversion Rate (Filter)

Press the MODE button. The MODE menu is displayed. Press the number “3” key followed by

ENTER. The CHANNEL SELECT menu is displayed. Choose the channel whose signal is to be
filtered using the number keys. For example, “1” + ENTER. The SETUP/CAL menu is displayed.




Note: Formulae too long to be completely displayed will display the Hastings
Instruments Gas ID number. Choosing “0” will blank the GASID elements on the

display.