Page 40 of 42


Congratulations!

You have just calibrated your Power

Pod

-400 to read zero at zero volts and the transducer’s max

display value at the transducer’s max signal input.


The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly

to the Power

Pod

-400.


*

Individual channels may be calibrated with calibrated transducers connected, breakout

connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not
supplied). Such procedures are not covered in this document.

!

If the calibration for the channel just zeroed had been severely compromised prior to beginning

this procedure, zero may not be immediately displayed and you may have to repeat the procedure

after resetting the span. Check that the span has been set correctly!

#

If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply

may need to be repaired.