Fluke 35040 User Manual

Page 20

Advertising
background image

35040
Operators Manual

2-6

Figure 2-2. Leakage Current and Start of Exposure

When automatic exposure detection is disabled (start and stop threshold set to 0.0 pA), there is no
correction for extrapolated charge due to leakage current. The dosimeter must be manually reset
between exposures and the leakage current kept as low as possible for best results.

Exposure Charge and Dose

The charge on the electrometer during an exposure is due to the incident radiation plus the charge that
was on the electrometer before the exposure plus the extrapolated charge due to any leakage currents in
the ion chamber, cables, or electrometer. To obtain only the charge due to the exposure, the instrument
employs the following equation:



The exposure charge variable Q

Exposure

is set to zero at the start of an exposure either manually by

pressing the RESET/MEASURE button or automatically if the start and stop thresholds are used. Over
the duration of the exposure, Q

Exposure

and Dose are updated about once a second. The displayed

measurement value (4-1/2 digit) is held following the exposure.

Timed Charge and Dose

This measurement accumulates charge only while the elapsed time is less than or equal to the timer
interval setting. The displayed value is held following the completion of the timer interval.

(

)

(

)

[s]

[s]

[A]

[C]

[C]

1

2

Leakage

1

2

Timed

T

T

I

Q

Q

Q

×

=

ADCF

ICCF

Q

Dose

Timed

Timed

Ч

Ч

=

[Units/C]

[C]

Time

0

- Charge samples in exposure.

- Extrapolated charge due to leakage current

QOffset

- Charge samples in Leakage.

Charge

- Samples discarded.

TOffset

Start of Exposure

TStart

X X

X

X

ADCF

ICCF

Q

Dose

Exposure

Ч

Ч

=

[Units/C]

[C]

(

)

(

)

[s]

[s]

[A]

[C]

[C]

Offset

Leakage

Offset

Exposure

T

T

I

Q

Q

Q

Ч

=