Troubleshooting and Service

Low Battery

4


4-7

4.4 Low Battery

Go to

line:

1.

Is the instrument connected to line voltage? Is the instrument connected to line
voltage?

YES

NO

2
6

2.

Turn the instrument off. Wait a few seconds. Turn the instrument on. A
sequence of screens will appear. On the last screen, the battery voltage and
current are displayed. Is the battery current between +0.10 A and +3.00 A?

YES

NO

3
8

3.

The battery is charging. Has the instrument been plugged in for more than 8
hours?

YES

NO

4

--

4.

Is the battery voltage greater than 5.9 volts?

YES

NO

5

--

5.

PROBLEM: False battery low indication. See Section 4.10.


6.

Connect AC line cord. If the display goes blank or stops updating, turn the
instrument off. Wait a few seconds. Turn the instrument on. Is the battery
current between +0.10 A and +3.00 A?

YES

NO

7
8

7.

The battery is charging. "LoBat" indicator will turn off when the battery voltage
is above 5.9 volts.

8.

Is the battery current between -1.00 A and -0.10 A?

YES

NO

9

18

9.

Disconnect AC line cord from rear panel. Remove line fuses from rear panel
fuse holder. Test the fuses with an ohmmeter. Are the fuses open circuited?

YES

NO

10
11

10. See

Section

4.6.


11.

Reinstall fuses. Reconnect AC line cord.

Is the battery current between

+0.10 A and +3.00 A?

YES

NO

12
13

12.

The battery is charging.


13.

Attach known operational appliance to outlet. Does the appliance work?

YES

NO

14
16

14.

Instrument or line cord is faulty. Replace cord with known good cord or
establish that the line cord is good. Is line cord good?

YES

NO

15
17

15.

PROBLEM: Battery charger inoperative. See Section 4.10.


16.

Outlet is faulty.


17.

See Section 4.9. Order proper line cord.


18.

PROBLEM: Low battery during trickle charge. Call Fluke Biomedical for
Service.