Fluke 35040 User Manual

Page 40

Advertising
background image

35040
Operators Manual


3-16








Figure 3-23.

The Bias Select Screen With the Bias Set to a Positive Voltage








Figure 3-24.

The Bias Select Screen With the Bias Turned Off



Figure 3-25.

The Bias Select Screen With the Bias Set to a Negative Voltage


In the example of Figures 3-23 to 3-25, the maximum voltage setting is 300.0 V. Thus, 300.0 V is 100%
of 300.0 V and -150.0 V is -50% of 300.0 V.

Use the UP and DOWN buttons to scroll through the customized bias settings. When the bias setting is
selected, wait for the bias voltage to match the selection before pressing the RESET/MEASURE button.
The top line of the display will read “WAIT Bias Not Ready” until the measured voltage matches the
selection. If the RESET/MEASURE button is pressed while the bias voltage is changing, the bias may be
set to the wrong voltage. While the bias voltage is changing, a current is produced that may cause an
“Internal Reset Error”, see Section 3.9.12.

Changing the bias setting automatically zeroes all
measurement readings.


To prevent an electrical shock, the Model 35040
Therapy Dosimeter bias should always be set to 0
volts or the instrument turned off when connecting
or disconnecting an ion chamber or triaxial cable.

Set:

300.0 V 100%

Central Electrde Pos

IC Bias: 300.0 V CEP

ADC

Set:

0.0 V 0%

Bias Off

IC Bias: 0.0 V OFF

ADC

Set:

-150.0 V -50%

Central Electrde Neg

IC Bias:-150.0 V CEN

ADC

NOTE

WARNING