35040
Operators Manual


4-8

4.5 Bias Error

Go to

line:

1.

Is "??Bias" displayed?

YES

NO

2
6

2.

Has the instrument been on for more than one minute?

YES

NO

3
8

3.

Turn off the power. Wait 10 seconds. Turn on the power. Wait 1 minute. Is
"??Bias" displayed?

YES

NO

4

--

4.

PROBLEM: Electrometer fault. See Section 4.10.

YES

NO

5

--

5.

Change the bias voltage to a different setting and then change back. Wait
about 10 seconds. Is the error on the display?

6.

Change the bias voltage to 0.0 V. Disconnect all triax cables. Cap the
connectors. Change bias voltage to desired setting. Wait about 10 seconds.
Is the error on the display?

YES

NO

7
8

7.

Change the bias voltage to 0.0 V. Examine the triax connectors on the
instrument, both front and rear, for foreign matter. Is foreign matter present?

8. See

Section

4.7.

YES

NO

9

18

9. Normal

operation

established.

YES

NO

10
11

10.

Do you use a separate cable to connect the ion chamber to the instrument?


11.

Change the bias voltage to 0.0 V. Connect only the cable to the instrument.
Change the bias voltage to the desired setting. Wait about 10 seconds. Is the
error on the display?

YES

NO

12
13

12.

PROBLEM: Bias fault in cable. See Section 4.10.


13.

Change the bias voltage to 0.0 V. Connect the ion chamber. Change the bias
voltage to the desired setting. Wait about 10 seconds. Is the error on the
display?

YES

NO

14
16

14.

PROBLEM: Bias fault in instrument. See Section 4.10.

YES

NO

15
17

15.

PROBLEM: Bias fault in ion chamber. See Section 4.10.


16.

This message is displayed until the signals from the isolated side are valid.