B.4.6 Annunciators, Indicators, and Overrange Conditions

Table B-1 specifies the annunciators that may appear on the top line of the display.

Table B-1. Annunciators

Annunciator Condition

??Bias

No bias measurement possible since the
electrometer is not ready

ADC

Air density correction in use (only applicable to
radiological measurements)

ALC

Automatic leakage compensation enabled

LoBat

Low Battery voltage

LoBias

Low Bias voltage

HiBias High

Bias

voltage

OvAmps

Overranged current

Rst Resetting

electrometer

Tmr Timer

in

use

WAIT Bias Not Ready

The dosimeter is setting the bias

WAIT System Not Rdy

The dosimeter is resetting the electrometer
(only applicable when automatic leakage
compensation is enabled)


Figure B-8 illustrates the locations of some annunciators. The battery, bias, and overranged current
annunciators appear at the same location, and the instrument alternates the displaying of each
annunciator when more than one is applicable.

Figure B-8.

Annunciator Positions


Table B-2 lists the indicators the instrument uses to signal that a measurement quantity has not been
compensated for leakage or is not available due to the specified condition.

Table B-2. Measurement Indicators and Conditions

Indicator Measurement

Condition

? (left of value) All

Non-leakage

1

Automatic leakage compensation is enabled, but the exposure
started before a leakage measurement was available.

-.---

Leakage

Automatic leakage compensation is enabled, but the exposure
started before a leakage measurement was available.

INVALID

Effective Time

The calculated effective time is longer than the actual
exposure. This condition occurs when a large amount of
charge is injected at the beginning of the exposure.

NA

Leakage

The leakage measurement is not available since automatic
leakage compensation is disabled.

OVER Non-leakage

1

The quantity or one of its dependents has exceeded the
maximum value. Refer to Table B-3 for more information.

1

All non-leakage measurements (e.g., charge, current, dose, rate, effective time, etc.)

3 9 . 9 9 5 n C

5 9 . 8 9 1 s

6 5 9 . 8 p A

A D C L o B i a s T m r A L C