Fluke 35040 User Manual

Page 44

Advertising
background image

35040
Operators Manual


3-20






Figure 3-32.

Annunciator positions


Table 3-2 lists the indicators the instrument uses to signal that a measurement quantity has not been
compensated for leakage or is not available due to the specified condition.

Table 3-2. Measurement Indicators and Conditions

Indicator Measurement

Condition

? (left of value) All

Non-leakage

1

Automatic leakage compensation is enabled, but
the exposure started before a leakage
measurement was available.

-.--- Leakage

Automatic leakage compensation is enabled, but
the exposure started before a leakage
measurement was available.

INVALID Effective

Time

The calculated effective time is longer than the
actual exposure. This condition occurs when a
large amount of charge is injected at the
beginning of the exposure.

NA Leakage

The leakage measurement is not available since
automatic leakage compensation is disabled.

OVER Non-leakage

1

The quantity or one of its dependents has
exceeded the maximum value. Refer to Table 3-
3 for more information.

1

All non-leakage measurements (e.g., charge, current, dose, rate, effective time, etc.)


Table 3-3 specifies the conditions that cause the Model 35040 Therapy Dosimeter to display an
overrange indicator (OVER) for a measurement quantity.

Table 3-3. Overrange Conditions

Condition Overranged

Measurement

1

charge > 19.999 mC

• charge
• average current
• accumulated charge
• effective exposure time

accumulated charge > 1.0000 C

• accumulated charge

current > 1.000 μA

• current

current > 1.200 μA

• all non-leakage

time > 19,999.9 s

• effective exposure time

1

Includes radiological measurements derived from electrical measurements.

3 9 . 9 9 5 n C

5 9 . 8 9 1 s

6 5 9 . 8 p A

A D C L o B i a s T m r A L C