Fluke 35040 User Manual

Page 36

Advertising
background image

35040
Operators Manual


3-12

With the automatic leakage compensation settings displayed (Figure 3-14), press the TEST FUNCTION
button. The "f" cursor will appear to the left of the ALC annunciator. Use the UP and DOWN arrow
buttons to toggle the setting between on (ALC) and off (OFF).

With automatic leakage compensation enabled, press the TEST FUNCTION button to advance the "f"
cursor to the next field or exit Edit mode. When the cursor points to a threshold setting, use the UP and
DOWN arrow buttons to increment and decrement the value, respectively. Nominal values for the start
and stop thresholds are 1.3 and 1.2 pA, respectively. However, depending on the system leakage, other
values may be necessary.

When automatic leakage compensation is disabled, the annunciator displays off (OFF) (Figure 3-15), the
start and stop threshold settings cannot be changed and display Not Available (NA), as do the leakage
measurements.







Figure 3-14.

Automatic Leakage Compensation Enabled







Figure 3-15.

Automatic Leakage Compensation Disabled



The instrument may reset all measurements when
the automatic leakage compensation parameters
are
changed.


Disabling automatic leakage compensation prevents
automatic exposure detection and the measurement
of leakage current, and leakage measurements will
display Not Available (NA).


Figure 3-16 shows the name of the calibration agency and the date the instrument was calibrated. The
bottom three lines may be different on your instrument depending on what the calibration agency puts
here.







Figure 3-16.

The Test Function Screen 5 - Calibration Identification

Start Thres: 1.0 pA
Stop Thres: 0.9 pA
Current: 0.0 pA

ADC ALC

Start Thres: NA pA
Stop Thres: NA pA
Current: 0.0 pA

ADC OFF

NOTE

NOTE

Calibrated by

Keithley RMD

Nov 15, 1993