Fluke 35040 User Manual

Page 22

Advertising
background image

35040
Operators Manual

2-8

or dielectric absorption. The start and stop thresholds are set independently, but the stop threshold must
be less than the start threshold. The thresholds are settable in 0.1 pA increments from 0.0 pA to 9.9 pA.
In some test setups, it may be necessary to raise the thresholds to reliably detect an exposure in the
presence of a relatively large leakage current. Since the start threshold determines the minimum dose
rate, it may be necessary to lower the thresholds for some exposures.

The minimum dose rate is

where I[A] is the start threshold setting.

When the start and stop thresholds are set to 0.0 pA, the automatic detection of start and end of exposure
is disabled. This effectively means that the instrument is measuring as if an exposure has been running
continuously. In this mode of operation, leakage current is not measured, the average current is
measured from the last time RESET/MEASURE was pressed, and charge will accumulate due to leakage
current.


Thresholds other than those used by the calibration
agency may render inaccurate measurements.

2.3.2 Exposure Time

Historically, the determination of exposure time for

60

Co has involved multiple exposures to eliminate the

effect of the shutter speed. By using the total charge and the average current for the exposure, the
effective exposure time can be determined using a single exposure. If the average current varies by more
than 10% during the exposure, this measurement is considered invalid and will be replaced by “INVALID.”





The effective exposure time is derived from the total charge and the average current. In order to get
enough data points to accurately calculate the time, the exposure must be a minimum of 3 seconds. To
avoid artifacts at the start and end of exposure, the average current is calculated over the interval starting
1 1/3 seconds after the start is detected to 1 1/3 seconds before the end of exposure.













[s]

[Units/C]

[A]

=

lier

RateMultip

ADCF

ICCF

I

DoseRate

Ч

Ч

Ч

NOTE

⎟⎟

⎜⎜

=

[s]

[s]

[C]

[C]

1

2

1

2

T

T

Q

Q

I

Avg

⎟⎟

⎜⎜

=

[A]

[C]

Avg

I

Q

Time

Advertising