Chapter 2 Installation

Series 24-HP Instruction Manual

2-4

IM-24-HP

Wafer-Style Flow Meter Installation

Install the wafer-style meter between two conventional pipe flanges of the
same nominal size as the flow meter. If the process fluid is a liquid, make
sure the meter is located where the pipe is always full. This may require
locating the meter at a low point in the piping system. Note: Vortex flow
meters are not suitable for two-phase flows (i.e., liquid and gas mixtures).
For horizontal pipelines having a process temperature above 300° F,
mount the meter at a 45 or 90-degree angle to avoid overheating the elec-
tronics enclosure. To adjust the viewing angle of the enclosure or dis-
play/keypad, see page 2-17.

Figure 2-3. Wafer-Style Flow Meter Installation

When installing the meter make sure the section marked “inlet” is positioned
upstream of the outlet, facing the flow. This ensures that the sensor head is
positioned downstream of the vortex shedder bar and is correctly aligned to
the flow. Installing the meter opposite this direction will result in completely
inaccurate flow measurement. To install the meter:

1. Turn off the flow of process gas, liquid or steam. Verify that the line

is not pressurized. Confirm that the installation site meets the required
minimum upstream and downstream pipe diameters.

2. Insert the studs for the bottom side of the meter body between the pipe

flanges. Place the wafer-style meter body between the flanges with the
end stamped “inlet” facing flow. Center the meter body inside the di-
ameter with respect to the inside diameter of the adjoining piping.

3.

Position the gasket material between the mating surfaces. Make sure
both gaskets are smooth and even with no gasket material extending in-
to the flow profile. Obstructions in the pipeline will disturb the flow and
cause inaccurate measurements

4.

Place the remaining studs between the pipe flanges. Tighten the nuts in
the sequence shown in Figure 2-2. Check for leaks after tightening the
flange bolts

Caution!

When using toxic or cor-

rosive gases, purge the

line with inert gas for a

minimum of four hours

at full gas flow before

installing the flow meter.