Series 24-HP Instruction Manual

Appendix E MODBUS Commands

IM-24-HP

E-5

The following registers contain the display units strings:

Registers

Variable

Data type

Units

Function code

Addresses

32007-32012

Volume flow units

string

03, 04

2006-2011

32001-32006

Mass flow units

string

03, 04

2000-2005

32025-32030

Temperature units

string

03, 04

2024-2029

32019-32024

Pressure units

string

03, 04

2018-2023

32031-32036

Density units

string

03, 04

2030-2035

32013-32017

Energy flow units

string

03, 04

2012-2017


Function codes 03 (read holding registers) and 04 (read input regis-
ters) are the only codes supported for reading these registers, and
function codes for writing holding registers are not implemented.
We recommend that the floating point and long integer registers be
read in a single operation with the number of registers being a mul-
tiple of two. If these data are read in two separate operations, each
reading a single 16-bit register, then the value will likely be invalid.

The floating point registers with values in display units are scaled
to the same units as are displayed, but are instantaneous values
that are not smoothed. If display smoothing is enabled (non-zero
value entered in the Display TC item in the Display Menu), then
the register values will not agree exactly with the displayed val-
ues.

Exception Status Definitions

The Read Exception Status command (function code 07) returns the exception

status byte, which is defined as follows. This byte may be cleared by setting “coil”

register #00003 (function code 5, address 2, data = 0xff00).

Bit(s)

Definition

0-1

Byte order (see Modbus Order on page 2)

0 = 3-2:1-0 1 = 2-3:0-1

2 = 1-0:3-2 3 = 0-1:2-3

2

Temperature sensor fault

3

Pressure sensor fault

4

A/D converter fault

5

Period overflow

6

Pulse overflow

7

Configuration changed

Discrete Input Definitions