Chapter 2 Installation

Series 24-HP Instruction Manual

2-10

IM-24-HP

Installing Flow Meters with a Compression Connection*

Use the following formula to determine insertion length for flow meters
(NPT and flanged) with a compression process connection. The installa-
tion procedure is given on the next page.

Insertion Length Formula

I = S – F – R – t

Where:

I =

Insertion length.

S =

Stem length – the distance from the center of the sensor head to the base

of the enclosure adapter (S = 29.47 inches for standard probes; S = 13.1

inches for compact; S = 41.47 inches for 12-inch extended).

F =

Distance from the raised face of the flange or top of NPT stem housing to

the outside of the pipe wall.

R =

Pipe inside diameter

÷ 2 for pipes ten inches and smaller.

R =

Five inches for pipe diameters larger than ten inches.

t =

Thickness of the pipe wall. (Measure the disk cut-out from the tapping

procedure or check a piping handbook for thickness.)

Figure 2-6. Insertion Calculation (Compression Type)

Example:

To install a Series 241 meter with a standard probe (S = 29.47 inches) into a 14

inch schedule 40 pipe, the following measurements are taken:

F=3

inches

R=5

inches

t=0.438

inches

The insertion length for this example is 21.03 inches. Insert the stem through
the fitting until an insertion length of 21.03 inches is measured with a ruler.

*All dimensions are in inches