Series 24-HP Instruction Manual

Appendix E MODBUS Commands

IM-24-HP

E-9

The unit responds with an identical message to that transmitted, and the totalizer

is reset. If the “coil” is turned off as in the following message, the response is al-

so identical to the transmitted message, but the totalizer is not affected.

01 05 00 00 00 00 CD CA

01 Device address
05 Function code 5 = write single coil
00 00 Coil address = 0
00 00 Data to “turn off coil” does not reset totalizer
CD CA CRC