Lake Shore Model 642 Electromagnet Power Supply User’s Manual

Service

7-11

7.10 CONNECTOR

AND

CABLE DEFINITIONS

All non-power electrical connections to the rear of the Model 642 are detailed in this section.

7.10.1

Analog I/O Connector

The Analog I/O connector provides the connections for the External Programming voltage as well as Analog
representations of the current and voltage output levels. Although these inputs/outputs are electronically balanced to
minimize ground loops, the common-mode voltage should not exceed 5 V on the outputs and 2 V on the input.

1

8

9

15

Pin Name Pin Name

1
2
3
4
5
6
7
8

NC

Chassis-common

Current Program –

Chassis-common

Voltage Monitor –

Chassis-common

Current Monitor –

Chassis-common

9

10
11
12
13
14
15

NC

Chassis-common

Current Program +

Chassis-common

Voltage Monitor +

Chassis-common

Current Monitor +

Figure 7-7. ANALOG I/O Connector Details

7.10.2 Magnet

Connector

The Magnet Water connector provides the means to connect a water control valve (24 VAC) and an associated water
flow switch (closed during flow) to protect the magnet from loss of water flow. Pins 1 & 2 must be closed for normal
operation. Pins 3 & 4 supply 24 VAC at 1 A for operation of a water control valve.

4

1

Pin Name

1 Flow

Switch

(Com)

2 Flow

Switch

3

Magnet Water Valve A

4

Magnet Water Valve B

Figure 7-8. Magnet Connector Details