Lakeshore Learning Materials 642 User Manual

Page 14

Advertising
background image

Lake Shore Model 642 Electromagnet Power Supply User’s Manual

1-4

Introduction

Specifications (Continued)

Output programming

Internal current setting
Resolution

0.1 mA (20 bit)

Settling time

600 ms for 1% step to within 1 mA (of internal setting)

Accuracy

±10 mA ±0.05% of setting

Operation

Keypad, computer interface

Protection

Programmable current setting limit

Internal current ramp
Ramp rate

0.1 mA/s to 99.999 A/s (compliance limited)

Update rate

23.7 increments/s

Ramp segments

5

Operation

Keypad, computer interface

Protection

Programmable ramp rate limit

External current programming
Sensitivity

10 V/70 A

Resolution Analog
Accuracy

±10 mA ±1% of setting

Input resistance

20 k

Ω

Operation

Voltage program through rear panel, can be summed with internal current setting

Limits

Internally clamped at

±

10.1 V and bandwidth limited at 40 Hz to protect output

Connector Shared

15-pin

D-sub

Readings

Output current
Resolution 0.1

mA

Accuracy

±

5 mA

±

0.05% of rdg

Update rate

2.5 rdg/s display, 10 rdg/s interface

Output voltage (at supply terminals)
Resolution 1mV
Accuracy

±

5 mV

±

0.05% of rdg

Update rate

2.5 rdg/s display, 5 rdg/s interface

Front panel
Display type

8-line by 40-character graphic vacuum fluorescent display module

Display readings

Output current, output voltage, and internal water temperature

Display settings

Output current and ramp rate

Display annunciators

Status and errors

LED annunciators

Fault, Compliance, Power Limit, Ramping, Remote

Audible annunciator

Errors and faults

Keypad type

26 full travel keys

Keypad functions

Direct access to common operations, menu-driven setup

Power

White flush ON and black extended OFF push buttons

Advertising