Lake Shore Model 642 Electromagnet Power Supply User’s Manual

4-12

Operation


Table 4-3. Default Parameter Values

Output Settings

Output Current* .....................................0 A
Current Ramp Rate..................... 99.999 A/s

Maximum Settings

Max Output Current..........................70.1 A
Max Ramp Rate ......................... 99.999 A/s

External Program Mode

External Program Mode .................. Internal

Ramp Segments

Ramp Segments .............................Disabled
Ramp Segments Current ........................0 A
Ramp Segments Rate ................. 1.0000 A/s

Display

Brightness ..............................................25%

Keypad Locking

State............................................... Unlocked
Lock Code ............................................... 123

Computer Interface

Baud...................................................... 9600
IEEE-488 Address ....................................12
IEEE-488 Terminators.......................CR/LF
Mode*..................................................Local

Water Settings

Magnet Water................................ Disabled
Internal Water ............................... Disabled


* Indicates value is also initialized on power up