Lake Shore Model 642 Electromagnet Power Supply User’s Manual

Computer Interface Operation

5-37

STOP

Stop Output Current Ramp Command

Input: STOP

[term]

Remarks:

This command will stop the output current ramp within two seconds of sending the command.
To restart the ramp, use the SETI command to set a new output current setpoint.

XPGM

External Program Mode Command

Input: XPGM

<mode>[term]

Format:

n
<mode>

0 = Internal, 1 = External, 2 = Sum.

Example:

XPGM 1[term] – Places the Model 642 into external program mode where the output current is set by
an external voltage.

XPGM?

External Program Mode Query

Input: XPGM?[term]
Returned:
<mode>[term]
Format:

n

(Refer to command for description)