Lake Shore Model 642 Electromagnet Power Supply User’s Manual

3-6

Installation

POWER CABLE

STRIP 13mm (0.5 in)

WIRE FERRULE (OPTIONAL)

POWER TERMINALS

140 mm (5.5 in)

Figure 3-6. Typical Power Input Wiring

3.3.6 Wiring

Cover

When the power wiring, voltage setting and current setting are complete, install the wiring cover with the six (6)
6-32 × 3/8 screws provided. The wiring cover installation is shown in Figure 3.7. Be sure that the voltage indicator tab
of the voltage locator card (Figure 3.2) shows through the correct indicator slot in the wiring cover.

WARNING: Do not connect power or attempt to operate the unit with this cover removed. Lethal voltage and

currents exist inside. There is a risk of injury or death if an operator or technician comes in contact
with these potentials.

CHASSIS

WIRING COVER

6-32 X 3/8 SCREW

(6)

POWER CORD

VOLTAGE INDICATOR SLOTS

Figure 3-7. Wiring Cover Installation