TROUBLESHOOTING & REPAIR

F-36

F-36

POWER WAVE 355/405

AUXILIARY TRANSFORMER TEST (continued)

TEST PROCEDURE

1. Remove input power to the POWER WAVE

355/405.

2. Using a 5/16” nut driver, remove the case

wraparound cover.

3. Perform the Input Capacitor Discharge

Procedure detailed earlier in this section.

4.

Locate the auxiliary transformer. See Figure
F.10.

5.

Locate the secondary leads and plug P52.
See Figure F.10 and F.11.

FIGURE F.10 Auxiliary Transformer

FIGURE F.11 Plug Lead Connections Viewed From Transformer Lead Side of Plug

W

AR

NIN

G

RE

M

OT

E

PO

W

ER

OF

F

ON

Auxiliary
Transformer


Secondary Lead
Plugs P52

STATUS

THERMAL

LINCOLN

ELECTRIC

Com 2

532

PW405

Only

(220V)

(115V)

(31)

Plug P52