Procom VOR/ILS Analyzer EVS200 User Manual

Page 45

Advertising
background image

VOR/ILS Analyzer EVS200

0796.1800.02

E-8

36

Chapter 3: Operation

Command DD1

command

function

DD1

<DDM information>

query of the DDM value in

µµµµ

A

The command DD1 reads the actual DDM value in µA in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

DD1 <CR>
dd1 <CR>

e.g.

194.32uA <CR/LF>

e.g.

194.32uA <CR/LF>

Command SD0

command

function

SD0

<SDM information>

SDM value query

The command SD0 reads the actual SDM value (value without dimension) in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

SD0 <CR>
sd0 <CR>

e.g.

0.8006 <CR/LF>

e.g.

0.8006 <CR/LF>

Command SD1

Command

function

SD1

<SDM information>

query of SDM value in

µµµµ

A

The command SD1 reads the actual DDM value in µA in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

SD1 <CR>
sd1 <CR>

e.g.

774.87uA <CR/LF>

e.g.

774.87uA <CR/LF>