Beacon-mode – Procom VOR/ILS Analyzer EVS200 User Manual
Page 59
VOR/ILS Analyzer EVS200
0796.1800.02
E-8
50
Chapter 3: Operation
Beacon-Mode
Command AM4
command
function
AM4
<AM-modulation depth (F1-range)>
query of AM-modulation depth (F1)
The command AM4 reads the actual AM-modulation depth (F1 3000 Hz) in the BEACON mode.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
AM4 <CR>
am4 <CR>
e.g.
29.3% <CR/LF>
e.g.
29.3% <CR/LF>
Command AM5
command
function
AM5
<AM-modulation depth (F2-range)>
query of AM-modulation depth (F2)
The command AM5 reads the actual AM-modulation depth (F2 1300 Hz) in the BEACON mode.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
AM5 <CR>
am5 <CR>
e.g.
30.5% <CR/LF>
e.g.
30.5% <CR/LF>
Command AM6
command
function
AM6
<AM-modulation depth (F3-range)>
query of AM-modulation depth (F3)
The command AM6 reads the actual AM-modulation depth (F3 400 Hz) in the BEACON mode.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
AM6 <CR>
am6 <CR>
e.g.
29.8% <CR/LF>
e.g.
29.8% <CR/LF>