Humidifiers

A humidifier may be required if you are experiencing dryness of the nose, throat,
or mouth. The following ResMed humidifiers are compatible with the
VPAP Adapt SV:

Figure 7: Humidifiers compatible with the VPAP Adapt SV

When using a humidifier, stop the VPAP Adapt SV by using the Start/Stop key or
by removing the mask to trigger the SmartStop function. Do not switch the
VPAP Adapt SV off at the power switch or the power source while it is running, as
this could result in condensation in the motor.

Please refer to the relevant humidifier user manual for detailed instructions and
warnings.

!

CAUTIONS

Be very careful not to allow water to enter the VPAP Adapt SV
as this could affect therapy and/or damage the device.

Do not tip the humidifier while it is connected to the
VPAP Adapt SV
. If water enters the VPAP Adapt SV device, turn off the
main power switch at the back of the unit and unplug it from the power
outlet. Return the unit to your equipment supplier for checking.

Connecting a Humidifier

Note: After connecting or removing a humidifier, always perform the Air Circuit
Learn function (see page 21).

!

WARNING

When using a humidifier, position it so that it is lower than your head will
be during sleep, and at the same level or lower than the VPAP Adapt SV.

HumidAire 2i

heated humidifier

Medium air tubing

21" (52 cm)

Humidifier accessory (HumidAire only)

HumidAire

heated humidifier