Alarm specifications, Iec 601-1 classifications – ResMed Flow Generator VPAP Adapt SV User Manual

Page 34

Advertising
background image

34

Alarm specifications

High maximum mask pressure alarm
Stops air delivery at 30 cm H

2

O >700 msec.

High mean mask pressure alarm
Stops air delivery when mean mask pressure exceeds 22 cm H

2

O (averaged

over approx 1 minute).

Low mask pressure alarm
Mask pressure is less than 3 cm H

2

O for longer than 5 seconds while the

patient is using the VPAP Adapt SV.

Insufficient pressure support alarm

Is triggered if for 3 breaths in a row the maximum mask pressure achieved at
inspiratory to expiratory transition is either <87.5% of the absolute mask
pressure targeted at that point, or < absolute mask pressure targeted minus
1 cm H

2

O, whichever is the lower.

High leak alarm
Leak exceeds 30 L/min for more than 20 seconds.

Flow blocked alarm
Flow is less than 6 L/min for more than 20 seconds.

IEC 601-1 classifications

Class II, Type CF

Class II medical electrical equipment provides protection against electric shock
either by double insulation or reinforced insulation and does not require a
protective earth.

Inspiratory Trigger Characteristics

The inspiratory trigger occurs at a flow greater than 0 LPM when tested with
EPAP=5 cm H

2

O, IPAP=15 cm H

2

O, sinisoidal breath shape. The inspiratory

trigger response time is <600 ms when tested with EPAP=5 cm H

2

O,

IPAP=15 cm H

2

O, 0.67 L/sec inspiratory and expiratory flow rate. The time is

measured from the beginning of inspiratory flow to the beginning of pressure
rise to IPAP.

Note: The ASV algorithm is designed to treat a patient with essentially normal
lungs. Hence, trigger response is less sensitive and pressure ramp-up is more
gentle than typical non-invasive ventilators.

228234 VPAP Adapt SV User Guide USA Eng.book Page 34 Monday, October 13, 2008 2:15 PM

Advertising