Technical Specifications

English

Displayed values

Maximum low impedance flow at stated pressures

Glossary of symbols

Read the relevant operating manual before using

Type CF equipment

AC switch

Class II

Drip proof

Value

Range

Accuracy

Display
Resolution

Proximal cuff

Pressure

-3 to 40 cm H

2

O

0.5 cm H

2

O

0.1 cm H

2

O

Mass flow sensor

Leak

0–99 liters per minute

+/- 12 L/min

1 liter per breath

Respiratory rate

8–30 breaths per minute

+/- 10%

1 breath per minute

Tidal volume

0–4500 milliliters

+/- 10%

1 milliliter

Minute ventilation

0–40 liters per minute

+/- 10%

0.1 liters per minute

Target MV

0–40 liters per minute

+/- 10%

0.1 liters per minute

Pressure (cm H

2

O)

Flow (L/min)

4.2

217

8.6

218

13.0

219

IPX1