Air Filter

Inspect the air filter every month and check if it is blocked by dirt or contains
holes.

The filter should be replaced every six months, or more often if the flow
generator is operating in a dusty environment.

!

WARNING

Do not wash the air filter once it has become soiled. The air filter is not
washable or reusable.

Fitting the filter

Remove the filter cover at the back of the flow generator.

Remove and discard the old air filter.

Insert a new filter, the blue tinted side facing towards you.

Note: Match the cut corner of the filter with the shape of the casing.

Replace the air filter cover. The blue tinted side of the filter should be facing
out.

Hypoallergenic Air Filter

The ResMed hypo-allergenic filter is a standard filter to which an electrostatic
filter has been bonded to capture very small particles. The filter is intended for
those users of ResMed flow generators who will benefit from enhanced filtering
of the air delivered during treatment.

Inspect the air filter every month and check if it is blocked by dirt or contains
holes.

The filter should be replaced every three months, or more often if the flow
generator is operating in a dusty environment.

Fit as described above in “Air Filter”.

Servicing

This product (VPAP Adapt SV) should be inspected by an authorised ResMed
Service Centre five years from the date of manufacture. Prior to this, the device
is intended to provide safe and reliable operation provided that it is operated and
maintained in accordance with the instructions provided by ResMed.

Air filter cover