The VPAP Adapt SV unit is fitted with alarms to alert you to changes that will
affect your treatment. The most common reason for an alarm to sound is
because the system has not been properly assembled. Check that the air tubing
and pressure sensor tube have been properly attached to the flow generator and
mask (and humidifier if used).

You can mute an alarm by pressing the Alarm Mute key once. Unless the alarm
requires you to turn the power off, you can press any key again to remove the
alarm message from the LCD screen and to access all menus. If the problem is
still present, the alarm will sound and display again after one minute. The Alarm
LED will remain lit for as long as the problem is present.

We recommend that you test the audible alarm once a week. To do this, hold
down the Mute key while the VPAP Adapt is starting up. If the alarm is
functioning correctly it will beep twice.

Low Mask Pressure Alarm

Triggered by:

pressure sensor line is blocked or disconnected

air pressure at the mask has fallen below a set level

mask is removed and SmartStop has been disabled.

If a low pressure alarm sounds:

1. Check that the air tube and sensor tube are connected properly.

2. Turn the VPAP Adapt SV off and on again at the power switch. If the alarm

persists, return the unit to ResMed for servicing.

LCD message

Alarm type

Action

LOW PRESSURE!
Check circuit

Low mask pressure
alarm

Check tube connections.

LCD turns off

Power fail alarm

Turn off VPAP Adapt SV at power switch.

HIGH PRESSURE!
Turn power off

High pressure alarm

Turn off VPAP Adapt SV at power switch.

HIGH AVG PRESS!

High pressure alarm

Turn off VPAP Adapt SV at power switch.

HIGH LEAK!
Check circuit

High mask leak alert

Adjust the mask.

LOW PS!
Check circuit

Insufficient pressure
Support alarm

Check tubing.
Run LEARN CIRCUIT.

FLOW BLOCKED!
Turn power off

Flow blocked alarm

Turn off VPAP Adapt SV. Check air
circuit. Turn on VPAP Adapt SV.