Humidaire – ResMed Flow Generator VPAP Adapt SV User Manual

Page 12

Advertising
background image

12

HumidAire™

You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV
unit to the HumidAire™ (see Figure 7).

1

Make sure both the HumidAire and the VPAP Adapt SV are turned off.

Fill the HumidAire with water as described in the humidifier manual. Place the
filled water chamber inside the HumidAire.

2

Gently detach enough of the tube clips to release the end of the pressure sensor
tube from the VPAP Adapt SV air tubing.

Connect the VPAP Adapt SV air tubing to the left connector port on the humidifier,
and the medium 21" (52 cm) air tubing to the right connector port. Close the
HumidAire lid.

3

Place the VPAP Adapt SV on top of the HumidAire. Never place the
VPAP Adapt SV unit underneath the humidifier. (This is to avoid water spilling into
the unit.)

4

Connect the free end of the medium air tubing to the VPAP Adapt SV.

Connect the pressure sensor tube by twisting the Luer lock onto the
VPAP Adapt SV (see page 9).

HumidAire

Medium air tubing

VPAP Adapt SV air tubing

Pressure sensor tube

228234 VPAP Adapt SV User Guide USA Eng.book Page 12 Monday, October 13, 2008 2:15 PM