Omega Speaker Systems DRX-TC User Manual

Page 15

Advertising
background image

Note 1:

Bit 2&3 set the speed of process totalize. The time it would take process totalize to
reach the non-totalize reading is shown in the following table:

Bit 3

Bit 2

Time to reach regular reading

0

0

1 minute

0

1

1 hour

1

0

1 day

1

1

30 day

3. DECIMAL POINT :

This is one byte (2 characters) as:

value

meaning

1

XXXXXX.

2 XXXXX.X
3

XXXX.XX

4

XXX.XXX

5

XX.XXXX

6

X.XXXXX

Note 1:

For TC and RTD models, value must be 1 to 3 only.

4. FILTER TIME CONSTANT:

This is one byte (2 characters) as:

value

meaning (per no. of readings)

0

No filtering

1

2

2

4

3

8

4

16

5

32

6

64

7

128