DRX-ACV SPECIFICATIONS

INPUT RANGES:
400 mV, 4V, 40 Volt, 400 Volt

FREQUENCY RANGE:
30 Hz to 1 KHz

INPUT IMPEDANCE:
2.1 Meg for all ranges.

ISOLATION:
Dielectric strength to 1000 Vrms transient per
1 minute test based on EN 61010 for 50 Vdc or Vrms
working voltage.
Three way Isolation:

• Power to input
• Power to analog output/communication
• Input to analog output/communication

INPUT OVER-VOLTAGE PROTECTION:

10% Above full scale continuously.
100% Above full scale for 10 seconds.

ANALOG TO DIGITAL TECHNIQUE:
Dual slope

READ RATE:
3 readings/second

ACCURACY AT 25 °C:
400 mV, 4V, 40V and 400V ranges
49 Hz to 500 Hz ±0.2% of FS
30 Hz to 1KHz ±0.2% of FS ±10 counts

TEMPERATURE STABILITY:
400 mV and 40 Volt range 150 ppm/°C typical
4V and 400 Volt range 100 ppm/°C typical

STEP RESPONSE FOR RS485 OUTPUT:
2 seconds to 99% of the final value
(Filter time constant = 64)

RESPONSE TIME:
To verify the response time, check the carriage return
<CR>, it will be sent at the end of the response. You
can send another command after you receive the
<CR>.
i.e.

send: *01X01
response: 01X01<DATA><CR>

Note: 01 is address.

WARM UP TO RATED ACCURACY:
30 minutes

INPUT POWER:
10 to 32 Volt DC

POWER CONSUMPTION:
2.4 watts (100 mA at 24 Vdc)

OPERATING AMBIENT:
-5 to +55 °C

STORAGE TEMPERATURE:
-40 to +85 °C

RELATIVE HUMIDITY:
90% at 40 °C non-condensing

SIG HI

N/C
N/C

SIG LO

+

+

-

N/C

SIGNAL INPUT

D

C

P

W

R

1 2