DRX-TC SPECIFICATIONS

INPUT TYPE:
Type J,K,T,E,R,S,B,N, J DIN
Input Type Range °F

Range °C

J

-346 to 1400°F

-210 to 760 °C

K

-454 to 2500°F

-270 to 1372 °C

T

-454 to 752 °F

-270 to 400 °C

E

-454 to 1832 °F

-270 to 1000 °C

R S

-58 to 3214 °F

-50 to 1768 °C

B

+212 to 3300 °F

+100 to 1820 °C

N

-454 to 2372°F

-270 to 1300 °C

J DIN

-328 to 1652 °F

-200 to 900 °C

ISOLATION:
Dielectric strength to 1000 Vrms transient per
1 minute test based on EN 61010 for 50 Vdc or Vrms
working voltage.
Three way Isolation:

• Power to input
• Power to analog output/communication
• Input to analog output/communication

INPUT OVER-VOLTAGE PROTECTION:

250 V AC for 1 minute

ANALOG TO DIGITAL TECHNIQUE:
Dual slope

READ RATE:
3 readings/second, automatic polarity

ACCURACY AT 25 °C:
±1°C for temperature above -150 °C
±2°C for temperature below -150 °C

TEMPERATURE STABILITY:
0.05 °C/°C

STEP RESPONSE FOR RS485 OUTPUT:
2 seconds to 99% of the final value
(Filter time constant = 64)

RESPONSE TIME:
To verify the response time, check the carriage return
<CR>, it will be sent at the end of the response. You
can send another command after you receive the
<CR>.
i.e.

send: *01X01
response: 01X01<DATA><CR>

Note: 01 is address.

WARM UP TO RATED ACCURACY:
30 minutes

INPUT POWER:
10 to 32 Volt DC

POWER CONSUMPTION:
2.4 Watts (100 mA at 24 V DC)

OPERATING AMBIENT:
-5 to +55 °C

STORAGE TEMPERATURE:
-40 to +85 °C

RELATIVE HUMIDITY:
90% at 40 °C non-condensing