11. Calibration

Your transmitter has been factory calibrated to meet or exceed the
specifications outlined in this manual. To maintain original
specifications it is generally recommended that your transmitter be re-
calibrated on an annual basis depending on operating conditions.

12. Calibration Procedure for HX93AC (all styles)

Relative Humidity Adjustment

Recommended equipment:
• Humidity Calibration Kit, OMEGA

®

Model No.: HX92-CAL

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

1. Remove enclosure cover
2. Connect transmitter as shown in figure below.
3. Apply power to transmitter and allow to warm up for 15 min.
4. Place unit in a 11% RH environment and allow to stabilize for 15 min.
5. Adjust potentiometer “P1” for a output reading of 5.76 mA
6. Place unit in a 75% RH environment and allow to stabilize for 15 min.
7. Adjust potentiometer “P2” for a output reading of 16.0 mA
8. Repeat steps 4, 5, 6, and 7 as necessary until proper readings are maintained.
9. Calibration complete.

Figure 6 - HX93AC Calibration Procedure

Relative Humidity Adjustment

16.00 mA

TB1

TB2

24V

PSU-93

HHM29

SENSOR HEAD

10A uAmA COM V

HX92-CA
CALIBRATION BOTTLE

P1 P3

P2

+

12

3

12

3

11

HX93A, HX93DA SERIES

RH/Temperature Transmitter

NOTE:
Do not adjust
P3. This is a
factory
setting.