14. Calibration Procedure for HX93AV (all styles)

Relative Humidity Adjustment

Recommended equipment:
• Humidity Calibration Kit, OMEGA

®

Model No.: HX92-CAL

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29B

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

1. Remove enclosure cover.
2. Connect transmitter as shown in figure below.
3. Apply power to transmitter and allow to warm up for 15 min.
4. Place unit in a 11% RH environment and allow to stabilize for 15 min.
5. Adjust potentiometer “P3” for a output reading of .110 Vdc.
6. Place unit in a 75% RH environment and allow to stabilize for 15 min.
7. Adjust potentiometer “P2” for a output reading of .750 Vdc.
8. Repeat steps 4, 5, 6, and 7 as necessary until proper readings are

maintained.

9. Calibration complete.

Figure 8 - HX93AV Calibration Procedure

Relative Humidity Adjustment

HHM29

SENSOR HEAD

HX92-CA
CALIBRATION BOTTLE

PSU-93

24V

.750 V

P3

TB1

P1

TB2

P2

+

uAmA

10A

COM V

13

2

3

2

1

14

HX93A, HX93DA SERIES

RH/Temperature Transmitter