Temperature Display Scaling

Recommended equipment:
• RTD Simulator OMEGA

®

Model No.: CL510-1

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

For ºF Display
1. Remove enclosure cover
2. Connect the power supply and multimeter as shown in the figure 7.
3. Disconnect the 2-wire Connector "J4" from the main Board.
4. Connect the RTD simulator to the "J4" header on the main board as

shown below.

5. Apply power to transmitter and allow to warm up for 10 min.
6. Set the RTD simulator for 0ºF output. (See figure on page 11.)
7. Adjust potentiometer "ZERO" on the back of the Temp display for a

reading of 0.0ºF

8. Set the RTD simulator for 150ºF output.
9. Adjust potentiometer "GAIN" on the back of the RH display for a

reading of 150.0ºF

10. Repeat steps 6, 7, 8 and 9 as necessary until proper readings are

maintained.

11. Calibration complete.
For ºC Display
12. Remove enclosure cover
13. Connect the power supply and multimeter as shown in the figure 7.
14. Disconnect the 2-wire Connector "J4" from the main Board.
15. Connect the RTD simulator to the "J4" header on the main board as

shown below.

16. Apply power to transmitter and allow to warm up for 10 min.
17. Set the RTD simulator for 0ºC output. (See figure on page 11.)
18. Adjust potentiometer "ZERO" on the back of the Temp display for a

reading of 0.0ºC

19. Set the RTD simulator to the 150ºC output.
20. Adjust potentiometer "GAIN" on the back of the RH display for a

reading of 150.0ºF

21. Repeat steps 6, 7, 8 and 9 as necessary until proper readings are

maintained.

Calibration complete.