Calibration procedure for hx93av (all styles), Hx93a, hx93da series rh/temperature transmitter, Model no.: hhm29 • dc power supply, omega – Omega Speaker Systems RH/Temperature Transmitter HX93A User Manual

Page 14

Advertising
background image

10

15. Calibration Procedure for HX93AV (all Styles)

Temperature Adjustment

Recommended equipment:
• 100 Ohm RTD Simulator OMEGA

®

Model No.: CL510-1

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

1. Remove enclosure cover
2. Connect the power supply and multimeter as shown in the figure

below.

3. Disconnect the 2-wire Connector “J4” from the main Board.
4. Connect the RTD simulator to the “J4” header on the main board as

shown below.

5. Apply power to transmitter and allow to warm up for 10 min.
6. Set the RTD simulator to the 0ºF position.
7. Adjust potentiometer “P5” on the main board for a output reading

of 0.023 Vdc

8. Set the RTD simulator to the 150ºF position.
9. Adjust potentiometer “P4” for a output reading of 0.901 Vdc
10. Repeat steps 6, 7, 8 and 9 as necessary until proper readings are

maintained.

11. Calibration complete.

Figure 9 - HX93AV Calibration Procedure

Temperature Adjustment

.901 V

HHM29

10A

24V

PSU-93

+

CL-301

P4

J4

TB1

P5

TB2

COM

uAmA

V

SENSOR HEAD

13

2

3

2

1

15