13. Calibration Procedure for HX93AC (all styles)

Temperature Adjustment

Recommended equipment:
• RTD Simulator OMEGA

®

Model No.: CL510-1

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

1. Remove enclosure cover
2. Connect the power supply and multimeter as shown in the figure

below.

3. Disconnect the 2-wire Connector “J4” from the main Board.
4. Connect the RTD simulator to the “J4” header on the main board as

shown below.

5. Apply power to transmitter and allow to warm up for 10 min.
6. Set the RTD simulator to the 0ºF position.
7. Adjust potentiometer “P5” on the main board for a output reading

of 4.43 mA @ 0°F

8. Set the RTD simulator to the 150ºF position.
9. Adjust potentiometer “P4” for a output reading of 18.46 mA 150°F
10. Repeat steps 6, 7, 8 and 9 as necessary until proper readings are

maintained.

11. Calibration complete.

Figure 7 - HX93AC Calibration Procedure

Temperature Adjustment

SENSOR HEAD

PSU-93

+

24V

18.41 mA

P4

TB1

1

3

2

TB2

P5

3

2

1

HHM29

J4

CL-301

uAmA

10A

COM V

13

HX93A, HX93DA SERIES
RH/Temperature Transmitter