15a. display scaling for hx93dac (all styles), Hx93a, hx93da series rh/temperature transmitter, Model no.: hhm29 • dc power supply, omega – Omega Speaker Systems RH/Temperature Transmitter HX93A User Manual

Page 15

Advertising
background image

11

15A. Display Scaling for HX93DAC (all styles)

Your HX93DA transmitter comes with dual built-in LCD indicators for
displaying both Humidity and Temperature. Before the meter
adjustment or re-scaling can be done the transmitters current output
should be checked as per the calibration procedure in Section 12 and 13
to verify the display will be adjusted correctly and match the analog
output.
Relative Humidity Display Scaling

Recommended equipment:
• Humidity Calibration Kit, OMEGA

®

Model No.: HX92-CAL

• Handheld Digital Multi-meter, OMEGA

®

Model No.: HHM29

• Dc Power Supply, OMEGA

®

Model No.: PSU-93

1. Remove enclosure cover
2. Connect transmitter as shown in figure 6.
3. Apply power to transmitter and allow to warm up for 15 min.
4. Place unit in a 11% RH environment and allow to stabilize for 15 min.
5. Adjust potentiometer "ZERO" on the back of the RH display for a

reading of 11.0%

6. Place unit in a 75% RH environment and allow to stabilize for 15 min.
7. Adjust potentiometer "GAIN" on the back of the RH display for a

reading of 75.0%

8. Repeat steps 4, 5, 6, and 7 as necessary until proper readings are

maintained.

9. Re-scaling complete.

Figure 10 - Transmitter Lid Rear View

15A

HX93A, HX93DA SERIES

RH/Temperature Transmitter

Z

ZERO POTENTIOMETER

RELATIVE HUMIDITY
INDICATOR

TRANSMITTER LID REAR VIEW

TEMPERATURE
INDICATOR

GAIN POTENTIOMETER

ZERO POTENTIOMETER

GAIN POTENTIOMETER

Z

G

G

+

+

Advertising