Voltage transmitter wiring examples, Relative humidity output calculations, Hx93a, hx93da series rh/temperature transmitter – Omega Speaker Systems RH/Temperature Transmitter HX93A User Manual

Page 9

Advertising
background image

5

Models: HX93AV, HX93AV-D HX93AV-RP1 (Voltage Output)

TB1 - Relative Humidity

TB2 - Temperature

1. + Power Supply

1. + Power Supply

2. + Output

2. + Output

3. - Power Supply

3. - Power Supply

7. Voltage Transmitter Wiring Examples

For voltage output models (0 to 1 Vdc)

Figure 5 - Voltage Transmitter Wiring Example

8. Relative Humidity Output Calculations

To calculate % Relative Humidity by measuring the current or voltage
output use the following formulas.
For current output:
% RH = (Current measured in miliamps – 4) ÷ .16
EXAMPLE:

(11.04 mA – 4) ÷ .16 = 44% RH

For voltage output:
% RH = (Voltage measured in volts x 100)
EXAMPLE:

.44 x 100 = 44% RH

2

3

1

2

1

3

RELATIVE HUMIDITY OUTPUT

TB1

TEMPERATURE OUTPUT

TB2

DPi8 Panel Meter

45%RH

PSU-93

DPi8 Panel Meter

24V

23

°C

+

+

+

7

HX93A, HX93DA SERIES

RH/Temperature Transmitter

Advertising