Technical Specifications

The mask contains passive venting to protect against rebreathing. As a result of
manufacturing variations, the vent flow rate may vary.

Pressure-flow curve – Swift FX Nano/ Swift FX Nano for Her

Vent flo

w rat

e (L/min)

0

10

20

30

40

50

60

70

80

4

6

8

10

12

14

16

18

20

22

24

26

28

30

Mask pressure (cm H

2

O)

Pressure-flow curve – Mirage FX/ Mirage FX for Her/ SoftGel/ Activa LT/ Micro/ Activa

Vent flo

w rat

e (L/min)

0

10

20

30

40

50

60

4

6

8

10

12

14

16

18

20

Mask pressure (cm H

2

O)

Swift FX

Nano/

Swift FX

Nano

for Her

Mirage FX/

Mirage FX

for Her

SoftGel

Activa LT Micro

Activa*

Pressure

Flow (L/Min)

4 cm H

2

O

20

19

19

19

19

19

8 cm H

2

O

29

28

28

28

28

28

12 cm H

2

O

37

34

34

34

34

34

16 cm H

2

O

43

40

40

40

40

40

20 cm H

2

O

49

45

45

45

45

45

22.5 cm H

2

O 52

-

-

-

-

-

30 cm H

2

O

61

-

-

-

-

-

Deadspace: Empty volume of mask to swivel. The deadspace of the mask varies
according to cushion size.

125 mL
(Wide)

83 mL

145 mL
(Large)

185 mL
(Medium)

128 mL
(Extra
large)

311 mL
(Shallow)

Therapy pressure (cm H

2

O)

4 - 30

4 - 20

4 - 20

4 - 20

4 - 20

4 - 20

Resistance: Drop in pressure measured (nominal) (cm H

2

O)

at 50 L/min

0,5

0.25

0.3

0.1

0.1

0.1

at 100 L/min 2,0

1.1

0.9

0.7

0.7

0.5

Sound - DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO
4871. The A-weighted sound power level and the A-weighted sound pressure level of
the mask, at a distance of 1M, with uncertainty of 3 dBA, are shown.
Power Level 25 dBA

26 dBA

29 dBA

30 dBA

30 dBA

32 dBA

Pressure
level at 1m

17 dBA

18 dBA

22 dBA

22 dBA

22 dBA

24 dBA

Advertising